• Banner

Environment and Burn-in Temperature Chambers

Rugged, dependable, precision instruments

Delta 9000 SeriesTM Temperature Chambers offer a wide choice of temperature ranges and capacities in addition to smart electronics. Field proven by satisfied customers for over four decades, Delta chambers are quality instruments that provide service longevity and integrity unique in their field. Over the years, Cohu has added advanced features to the proven basic chamber design and has refined system components and programming capability to deliver outstanding quality and value.

Cohu Test Handler and Inspection Solutions

  • Key Features

    • Temperature expandable to -184°C, utilizing optional LN2 cooling
    • Standard chamber coolant of 1050 psi LCO2
    • 1-1/4” diameter access port on the left side (9059 and 9064)
    • Three 7/8” diameter access ports in the rear (9076)
    • Test volume 1/3 cubic foot to 5-3/4 cubic feet
    • -73°C to +315°C, precisely controlled by advanced electronic circuitry
    • Adjustable over-temperature protection device
  • Standard Specifications

    • Temperature Accuracy – set-point to actual ±1.0°C
    • Temperature Deviation – after stabilization ±0.1°C
    • Repeatability ±0.25°C
    • Line Voltage Stabilization ±0.25°C for 105-125VAC or 220-240VAC line voltage change
    • Fail-safe adjustable over-temperature
  • Temperature Controller

    • Allows a program to be repeated up to 9,999 times
    • Temperature, time and rate of change parameters
    • Soak time up to 100 hours
    • IEEE-488 and RS232 external interfaces
    • Push button convenience, backlit LCD
    • Battery-backed memory
    • Settable bus address and baud rates
  • Standard Temperature Chamber

    • Thermometer
    • CO2 hose/filter
    • 1050 psi CO2 valve
    • Instruction manual

9000 Series Test Chambers

SPECIFICATIONS
Test Volume 0.37 cubic feet (10.5 liters)
Internal Dimensions 8″H x 10″W x 8″D
(20.32 x 44.45 x 52.70 cm)
Overall Dimensions 12 ¾”H x 17 ½”W x 20 ¾”D
(32.38 x 44.45 x 52.70 cm)
Temperature Range -73° to +315°C or -100°F to +600°F
OPTION -184°C (-300°F) with optional LN2
Heat-up Rate 24°C / Minute
Cool-down Rate 32°C / Minute
LC02 Consumption 5.6 lbs/hr at -53°C (-65°F)
Air Circulation Vertical air flow pattern
Power Requirements 120 VAC, 50 or 60 Hz., single phase, 1800w
208 VAC, 50 or 60 Hz., three wire, single phase, 1800w
208 VAC, 50 or 60 Hz., four wire, single phase, 1800w
240 VAC, 50 or 60 Hz., single phase, 1800w
Net Wt. (Approx.) 47 lbs (21.3 Kg)
Shipping Wt. (Approx.) 60 lbs (27.2 Kg)
Door Options Available Blank with Tray;
Ten 3/8″ Dia Ports with Tray;
Window and Five 3/8″ Dia Ports – No Tray
SPECIFICATIONS
Test Volume 0.45 cubic feet (12.7 liters)
Internal Dimensions 7″H x 16″W x 7″D
(17.78 x 40.64 x 17.78 cm)
Overall Dimensions 12 ¼”H x 28 ¾”W x 20 ¾”D
(31.11 x 73.02 x 53.02 cm)
Temperature Range -73° to +315°C or -100°F to +600°F
OPTION -184°C (-300°F) with optional LN2
Heat-up Rate 20°C / Minute
Cool-down Rate 25°C / Minute
LC02 Consumption 6.5 lbs/hr at -53°C (-65°F)
Air Circulation Horizontal air flow pattern
Power Requirements 120 VAC, 50 or 60 Hz., single phase, 1800w
208 VAC, 50 or 60 Hz., three wire, single phase, 1800w
208 VAC, 50 or 60 Hz., four wire, single phase, 1800w
240 VAC, 50 or 60 Hz., single phase, 1800w
Net Wt. (Approx.) 55 lbs (24.9 Kg)
Shipping Wt. (Approx.) 70 lbs (31.8 Kg)
Door Options Available Blank with Tray;
Ten 3/8″ Dia Ports with Tray;
Window and Five 3/8″ Dia Ports with Tray
SPECIFICATIONS
Test Volume 0.92 cubic feet (26.05 liters)
Internal Dimensions 10″H x 16″W x 10″D
(25.40 x 40.64 x 25.40 cm)
Overall Dimensions 17 7/8″H x 27 ½”W x 24″D
(46.40 x 69.85 x 60.96 cm)
Temperature Range -73° to +315°C or -100°F to +600°F
OPTION -184°C (-300°F) with optional LN2
Heat-up Rate 14°C / Minute (Optional – 60°C / Minute)
Cool-down Rate 36°C / Minute
LC02 Consumption 7 lbs/hr at -53°C (-65°F)
Air Circulation Vertical air flow pattern
Power Requirements 120 VAC, 50 or 60 Hz., single phase, 1800w
208 VAC, 50 or 60 Hz., three wire, single phase, 1800w
208 VAC, 50 or 60 Hz., four wire, single phase, 1800w
240 VAC, 50 or 60 Hz., single phase, 1800w
Net Wt. (Approx.) 85 lbs (38.6 Kg)
Shipping Wt. (Approx.) 100 lbs (45.4 Kg)
Door Options Available Blank with Tray;
Ten 3/8″ Dia Ports with Tray;
Window and Five 3/8″ Dia Ports – No Tray
SPECIFICATIONS
Test Volume 1.85 cubic feet (52.4 liters)
Internal Dimensions 11″H x 19½”W x 15″D
(27.94 x 49.53 x 38.10 cm)
Overall Dimensions 17 ¼”H x 32″W x 25″D
(43.82 x 81.28 x 63.50 cm)
Temperature Range -73° to +315°C or -100°F to +600°F
OPTION -184°C (-300°F) with optional LN2
Heat-up Rate 10°C / Minute
Cool-down Rate 22°C / Minute
LC02 Consumption 9.62 lbs/hr at -53°C (-65°F)
Air Circulation Horizontal air flow pattern
Power Requirements 120 VAC, 50 or 60 Hz., single phase, 1800w
208 VAC, 50 or 60 Hz., three wire, single phase, 1800w
208 VAC, 50 or 60 Hz., four wire, single phase, 1800w
240 VAC, 50 or 60 Hz., single phase, 1800w
Net Wt. (Approx.) 98 lbs (44.5 Kg)
Shipping Wt. (Approx.) 105 lbs (47.6 Kg)
Door Options Available Blank – No Tray;
Window and Five 3/8″ Dia Ports – No Tray
SPECIFICATIONS
Test Volume 2.2 cubic feet (62.3 liters)
Internal Dimensions 12″H x 20″W x 16″D
(30.48 x 50.80 x 40.64 cm)
Overall Dimensions 21 ½”H x 32 7/8″W x 30″D
(54.61 x 83.66 x 76.20 cm)
Temperature Range -73° to +315°C or -100°F to +600°F
OPTION -184°C (-300°F) with optional LN2
Heat-up Rate 8°C / Minute
Cool-down Rate 20°C / Minute
LC02 Consumption 11.6 lbs/hr at -53°C (-65°F)
Air Circulation Vertical air flow pattern
Power Requirements 120 VAC, 50 or 60 Hz., single phase, 1800w
208 VAC, 50 or 60 Hz., three wire, single phase, 1800w
208 VAC, 50 or 60 Hz., four wire, single phase, 1800w
240 VAC, 50 or 60 Hz., single phase, 1800w
Net Wt. (Approx.) 115 lbs (52.2 Kg)
Shipping Wt. (Approx.) 235 lbs (106.6 Kg)
Door Options Available Blank – No Tray;
Window and Five 3/8″ Dia Ports – No Tray
SPECIFICATIONS
Test Volume 5.78 cubic feet (163.7 liters)
Internal Dimensions 20″H x 20″W x 25″D
(50.80 x 97.16 x 95.3 cm)
Overall Dimensions 36 ¼”H x 38 ¼”W x 37 ½”D
(92.08 x 97.16 x 95.3 cm)
Temperature Range -73° to +315°C or -100°F to +600°F
OPTION -184°C (-300°F) with optional LN2
Heat-up Rate 8°C / Minute
Cool-down Rate 12°C / Minute
LC02 Consumption 15 lbs/hr at -53°C (-65°F)
Air Circulation Horizontal air flow pattern
Power Requirements 208 VAC, 50 or 60 Hz., three wire, single phase, 4800w
208 VAC, 50 or 60 Hz., four wire, single phase, 4800w
240 VAC, 50 or 60 Hz., single phase, 4800w
Net Wt. (Approx.) 232 lbs (105.9 Kg)
Shipping Wt. (Approx.) 413 lbs (188.5 Kg)
Door Options Available Blank – No Tray;
Window Door

Monitoring of temperature excursions between two present limits is provided by the Temperature Monitor. If limits are exceeded, power to the monitored unit is removed, a warning lamp is illuminated and an audible alarm is activated. When temperature is again within limits, the monitor may be reset for continued operation. A combination pyrometer scale provides reading at a glance with both Fahrenheit and Centigrade ranges.

Expertise